Product Brand Yanling
Product No POWER-420B General generator
Product annex No
● Monopolar cutting:
(1)Standard mode and intense mode (including pure cut, mixed one, mixed two, mixed three)
(2)Standard mode: digestive endoscopy, laparoscopy, hysteroscopy, etc., as well as open operations which require less thermal damage (such as skull, maxillofacial surgery, plastic burns. Less sparks during cutting, less carbonization.)
(3)Intensive mode: power density feedback technology, low-power efficient cutting, general surgery, chest surgery, orthopedics, urology and other operations.
● Monopolar coagulation: Spray coagulation mode: used with argon knife. Soft mode: use under endoscopy, better effect on coagulation of parenchyma
● Bipolar electrocoagulation: Bipolar macro mode: use under endoscopy. Standard mode: use in neurosurgery and general surgery
● Quality neutral plate control system
Technical data:
| Output Power | Standard mode | Intense mode | |
| Monopolar pure cut | 350W | 350W | |
| onopolar mixed one | 250W | 250W | |
| Mixed two | 200W | 200W | |
| Mixed three | 120W | 120W | |
| Standard mode | Soft mode | Spray mode | |
| Monopolar coagulation | 120W | 120W | 100W |
| Standard mode | Macro mode | ||
| Bipolar coagulation | 50W | 80W |
Class I CF type, anti-defibrillation common equipment